Joerger Logo

MODEL QVSC

 

QUAD, 300 MHZ, VISUAL READOUT
COUNTER / SCALER

 

FEATURES:

  • FOUR, COMPLETELY INDEPENDENT, 8 DIGIT COUNTER / SCALERS
  • MODE SELECTION FOR EACH CHANNEL
  • 300MHZ COUNTING SPEED
  • COUNTER GATE TIMES, .01,.1, 1 AND 10 SEC. OR A SCALER MODE
    PER CHANNEL
  • CRYSTAL CONTROLLED TIME BASE
  • MANUAL OR REMOTE SCALER OPERATION
  • EASY READING .8" DISPLAYS
  • TEST MODE
  • 19" RACK MOUNTED INSTRUMENT

 

The JOERGER ENTERPRISES, INC. MODEL QVSC contains four completely independent, 300Mhz counters in a 19" rack mounted instrument. It features direct counting without the use of prescaling. Each channels output is on eight, .8" displays with leading edge suppression for easy reading. The operating modes are selected by switches which can chose counter modes with synchronous gate times of .01, .1, 1, or 10 seconds or a scaler mode. A ± .01% crystal oscillator is provided to guarantee high performance operation. A test mode is provided that tests each channel in its selected operating mode. During scaler operation all the LED’s are tested by displaying all "8’s" in the reset mode.

In the scaler mode the channels selected can be controlled either manually by a gate and reset switch, or remotely via rear panel signals. These control signals are all buffered so that the units can be daisy chained using one as a master and the other as slaves. An overflow LED is provided for each channel as well as rear panel overflow signals.

As an option, channel 4 can be operated in a preset mode. A rear panel rotary switch is provided to select the output of decades 2 through 8. When the selected decade overflows, a stop signal is generated which can be used to control the system. This feature does not effect the performance of channel 4 which may be used to its full capability.

 

SPECIFICATIONS:

INPUTS: (each channel)

Input Impedance 50 ohms +10% D.C. coupled.
Pulse Repetition Rate 300Mhz Typical, 250Mhz min.
Pulse Pair Resolution 4nsec
Sensitivity -500mv
Pulse Width 2nsec minimum

 

MODE CONTROL: (each channel)

Frequency Mode Gate selections of .01, .1, 1, and 10 seconds
Scaler Mode Manual or remote operation

CONTROL SIGNALS:

Start/Stop/Reset Inputs

Start, Stop and Reset inputs require either a positive 3.5 volt signal 1usec wide or a short to ground. The inputs are on the rear panel with each consisting of two BNC connectors that are bridged to facilitate daisy chaining units for common control.

Start/Stop/Reset Outputs

Start, Stop and Reset outputs are available at rear panel BNC's. A +4 volt signal is generated in response to either a manual or electrical input for each signal. A Stop signal is also generated in response to the preselect mode. This allows one scaler to control other scalers in the system.

Carry Out

Carry Outputs are available at rear panel NBC’s. They are negative going signals capable of generating -500mv into 50 ohms.

MANUAL CONTROLS:

Start-Stop

A three position switch is provided to start and stop all four channels.

Reset

A push-button is provided to reset all four channels. This switch is also used to test the display. When the switch is depressed all eights will appear on the display.

Test

A test switch is provided that injects test signals into all four channels after the counting gate.

PRESELECT MODE:

A preselect mode is available. It uses Channel 4 as the control channel. Any decade overflow from decade 2 through 8 can be selected by a rear panel switch. When the selected digit overflows a stop output signal is generated. If this signal is connected to the stop input, all four channels will be inhibited at that time. To disable, set to "OFF".

DISPLAYS:

Data

Displayed on eight .8" LED displays with leading zero suppression.

Gate

An LED is provided to indicate when a counting gate is open.

Overflow

An LED is provided to indicate when that channel has overflowed.

TEMPERATURE RANGE: 0° C to 50° C
POWER REQUIREMENTS: 105V to 125V A.C. 50-440HZ, 30W, 220V A.C., 50HZ (optional)
SIZE: 3 1/2" X 19" X 14" chassis.
CONNECTORS: BNC

 

JOERGER ENTERPRISES, INC 166 Laurel Road, East Northport, NY 11731 USA

1-631-757-6200 (Fax 1-631-757-6201) www.joergerinc.com